Infusion Pump Calibration: Flow Rate and Volume Testing Explained

How infusion and syringe pumps are tested for flow rate accuracy, delivered volume and occlusion alarms using an infusion device analyser, with setup tips and what a calibration record should include.

  • By the Shiwi engineering team
  • 7 min read
Blog banner with an ECG trace and the title Infusion Pump Calibration: Flow Rate and Volume Testing Explained

Infusion pump calibration checks that a pump delivers the flow rate and volume it is programmed to deliver, and that its occlusion and other alarms work at the right point. Engineers test this with an infusion device analyser, which measures the fluid the pump actually delivers over time and the pressure at which it alarms on occlusion. The same approach applies to volumetric infusion pumps and syringe pumps, with a few differences covered below.

This guide explains each test, how to read the results, and what your records should show.

Why do infusion pumps need regular performance testing?

Infusion pumps often deliver drugs where a small error in rate matters: inotropes, insulin, sedatives, heparin and paediatric doses. A pump can look normal and still under- or over-infuse because of a worn pumping mechanism, a damaged door or sensor, or a fault after being dropped.

The US FDA’s list of reported infusion pump problems includes alarms that fail to sound for occlusion or air in the line, occlusion alarms without a real blockage, battery failures, and broken components causing over- or under-infusion. The FDA also advises biomedical engineers to carry out preventive maintenance as recommended by the manufacturer and to keep a record of pump software versions.

What is tested during infusion pump calibration?

Test What it shows How it is done
Flow rate accuracy Whether average delivered flow matches the set rate Pump runs at set rates into the analyser; measured average flow is compared with the programmed rate
Volume delivered Whether the pump stops at the programmed volume A volume to be infused is set; the analyser measures the actual volume delivered
Occlusion pressure and alarm Whether the pump alarms when the line is blocked, and at what pressure Analyser closes the outlet and records pressure build-up until the pump alarms
Bolus Whether a programmed bolus is delivered accurately Bolus volume measured by the analyser
Other alarms and functions Whether safety features work Air-in-line, door open, end of infusion or near-empty syringe, low battery, keypad and display checks
Electrical safety Whether earth and leakage values are within limits Safety analyser, as for other mains-powered equipment

Flow rate accuracy

The pump is primed, connected to the analyser and run at one or more set rates. Choose rates that reflect clinical use: a low rate typical of drug infusions and a higher rate typical of fluids. The analyser reports average flow over the test, and the difference from the set rate is expressed as a percentage. The result is compared with the accuracy stated by the manufacturer in the service manual.

Fluke Biomedical’s white paper on testing infusion devices notes that the IEC reference method weighs delivered fluid on a beaker and scale over a 120-minute test, while graduated burettes with a stopwatch or automated analysers can also give accurate results. Burette methods are prone to errors in reading the meniscus and in timing, which is one reason analysers are preferred in the field.

Why short tests can mislead

Most pumps do not deliver fluid perfectly smoothly. Peristaltic and piston mechanisms deliver in small pulses, and many pumps take time to settle after starting. A test that is too short, especially at low rates, can show a large error even on a good pump, or hide a real one.

IEC 60601-2-24, the particular standard for infusion pumps and controllers, reports performance with a start-up graph and trumpet curves. An article in Patient Safety & Quality Healthcare explains that flow during the first two hours is plotted in the start-up graph, with the first hour treated as warm-up, and that the trumpet curve shows how much flow varies over short and longer averaging windows. The curve narrows as the window lengthens, because averaging smooths short-term variation.

For routine calibration you do not need to reproduce the full type test, but the test duration and volume should follow the manufacturer’s procedure, so that results are comparable from year to year.

Volume delivered

A volume to be infused (VTBI) is programmed and the pump is allowed to finish. The analyser compares the measured volume with the set volume, and the engineer checks that the end-of-infusion alarm or keep-vein-open behaviour works as described in the manual.

Occlusion pressure and alarm

The analyser closes the downstream line and measures the pressure as it builds. The engineer notes the pressure at which the pump alarms and checks it against the manufacturer’s specification for each occlusion setting, if the pump has more than one. The time taken to alarm at low flow rates is also worth recording, because pressure builds slowly when flow is low.

After the alarm, the engineer releases the pressure as the manual describes. A pump that releases a large bolus when an occlusion clears is a patient risk, so some procedures also measure the volume released.

How is a syringe pump tested differently?

Syringe pumps use the same analyser and the same tests, with a few additions:

  • Syringe selection: the pump calculates flow from the syringe brand and size set on it. Test with a syringe brand and size the pump is configured for, and the one used on the wards.
  • Syringe detection: check that the pump correctly detects syringe size, plunger engagement and clamp position, and alarms if the syringe is not seated.
  • Low rates: syringe pumps are often used at very low rates, where start-up delay and mechanical slack have the most effect. Allow enough test time.
  • Near-empty and empty alarms: confirm these trigger before the syringe is fully emptied.

How should the test be set up?

Most poor results come from the setup rather than the pump. The Fluke white paper recommends removing all air from the line before each flow test and keeping the tubing between pump and analyser fairly rigid. A short checklist:

  • use the administration set or syringe type specified for that pump, and a new one where possible;
  • use distilled or deionised water, or the fluid the analyser manual specifies;
  • prime fully and remove air bubbles from the set and the analyser connection;
  • keep the pump and analyser at the heights the analyser manual recommends;
  • let the pump run for its stated warm-up time before relying on the reading;
  • confirm the analyser itself is within its calibration date.

What should an infusion pump calibration record include?

  • pump make, model, serial number, asset number, location and software version;
  • administration set or syringe brand and size used;
  • analyser make, serial number and calibration due date;
  • each set rate, measured average flow and percentage error, with the tolerance applied;
  • programmed and measured volume;
  • occlusion alarm pressure for each setting tested;
  • results of air-in-line, door, end-of-infusion, battery and other alarm checks;
  • electrical safety results, overall result, engineer’s name, date and next due date.

Recording software version supports the FDA’s advice on tracking it across your inventory. For a wider view of how pump testing fits with other equipment, see our biomedical preventive maintenance schedule.

Safety note: This is general guidance. Testing must follow the manufacturer’s service manual and be done by qualified engineers with calibrated analysers. Never test a pump while it is connected to a patient.

How Shiwi calibrates infusion and syringe pumps

Shiwi Enterprises is a NABL-accredited calibration laboratory (ISO/IEC 17025) in Jaipur. We calibrate infusion pumps and syringe pumps on site across India, using a calibrated infusion device analyser for flow, volume and occlusion tests, along with electrical safety testing. Each pump receives its own calibration certificate for NABH and internal audits, and the same team handles repairs if a pump fails.

Schedule infusion pump calibration

If your infusion or syringe pumps are due for calibration, send us the number of pumps, their models and your location through our quote request form. We visit hospitals, labs and clinics anywhere in India, including small towns.

Frequently asked questions

What is checked during infusion pump calibration?

Flow rate accuracy at set rates, volume delivered against the programmed volume, occlusion alarm pressure, bolus delivery where used, other alarms such as air-in-line and door open, and electrical safety.

What equipment is used to calibrate an infusion pump?

An infusion device analyser measures the fluid the pump delivers over time and the pressure at which it alarms on occlusion. The IEC reference method uses a beaker and precision scale over a 120-minute test.

Is a syringe pump tested the same way as a volumetric infusion pump?

The same analyser and core tests are used, but syringe pumps also need checks of syringe size and brand settings, syringe detection and near-empty alarms, and enough test time at low rates.

Which standard covers infusion pump performance?

IEC 60601-2-24 sets particular requirements for the basic safety and essential performance of infusion pumps and controllers, and reports flow performance using start-up graphs and trumpet curves.