Calibration · Repair · Maintenance
EEG
Calibration, repair and preventive maintenance for EEG, carried out on-site at your hospital or lab, anywhere in India.
- NABL-accredited lab · ISO/IEC 17025
- On-site visits across India
EEG: overview
An EEG machine records the electrical activity of the brain through electrodes on the scalp. Neurologists use it to diagnose epilepsy and other conditions, reading very small signals of a few microvolts. The amplifiers, filters and timing have to be accurate or the recording can be misread.
Shiwi Enterprises calibrates, services and repairs EEG systems in neurology departments and diagnostic centres. Our engineers come to your site anywhere in India and test the system with a calibrated signal source.
What we check
Parameters covered for EEG
Electrical safety
Services
Services for EEG
Calibration for EEG
EEG calibration injects known signals into each channel and checks that the system records the right amplitude and frequency.
What we check
- Amplitude (sensitivity) on each channel
- Filter and frequency response
- Noise and channel balance
- Timebase
- Electrical safety: protective earth, insulation resistance, leakage current
You receive a calibration certificate with the measured readings for each unit. Plan a calibration once a year and after any repair to the amplifier or head box.
Service & maintenance for EEG
Common EEG problems are noisy channels, faulty electrode cables and head boxes, and software or PC connection faults. We trace the noise source, repair or replace the faulty part and retest every channel.
- Preventive maintenance
- Electrode cable and head box faults
- Noise and grounding faults
- PC and acquisition faults
- Recalibration after repair
We service equipment from any brand. Send us the make, model and number of units, and we will reply with a quote and a visit date.
How it works
How on-site calibration works
- 01
Send your equipment list
Share it through the form, by email or on WhatsApp. We confirm the scope and send a quote.
- 02
Fix a date
We agree a visit date with your team and plan the work around patient care.
- 03
Calibration and testing
Our engineers test each device against reference analysers and record the results.
- 04
Certificates for your records
You receive a calibration certificate for each instrument, ready for NABH and internal audits.
Get a quote for EEG
Tell us the make, model, quantity and city. We reply with a quote and a visit plan.