NABH Medical Equipment Calibration Requirements: A Checklist for Hospitals
A practical checklist for NABH coordinators and biomedical engineers covering equipment inventory, calibration schedules, certificates, electrical safety and the records assessors ask for.

NABH expects a hospital to show that its medical equipment is planned, maintained, safe and fit for use, and calibration records are a core part of that evidence. In practice, NABH calibration requirements come down to a complete equipment inventory, a documented schedule, a valid calibration certificate for each relevant device, electrical safety testing, and records that link all of these together. The checklist below helps you find gaps before the assessor does.
This article is written for NABH coordinators, quality managers and biomedical engineers preparing for a first assessment, a surveillance visit or a renewal.
Which NABH standard covers medical equipment?
The current hospital standard is the NABH Accreditation Standards for Hospitals, 6th Edition, effective from 1 January 2025. It has ten chapters. The one most relevant to biomedical teams is Facility Management and Safety (FMS), which deals with the safety of the facility and its infrastructure, including electrical safety. Related expectations appear elsewhere too, such as the quality assurance and safety programmes for laboratory and imaging services in the Access, Assessment and Continuity of Care (AAC) chapter.
Electrical safety and the safe, effective use of medical equipment both feature in the 6th edition, so equipment teams should expect assessors to ask about them.
The exact wording of each objective element matters during assessment, so read the FMS chapter of the current edition alongside this checklist.
What do assessors usually want to see?
Assessors look for a system, not a stack of certificates. They will typically ask how you know what equipment you have, how you decide when it is checked, who checks it, what happens when it fails, and where the evidence is. If any one of these links is missing, the certificates alone will not carry the case.
A good test is to pick any device at random, say a defibrillator in the emergency department, and see whether you can produce its asset record, last calibration certificate, electrical safety result, maintenance history and next due date within a few minutes.
NABH calibration checklist for hospitals
Use this as a working list. Tick each item only when the evidence exists and can be produced on request.
1. Equipment inventory
- A single, current list of all medical equipment across every department, including OT, ICU, NICU, emergency, wards, laboratory and imaging
- Each entry has a unique asset ID, make, model, serial number, location and date of installation
- Each device is classified by criticality or risk
- Equipment on loan, on demonstration or under a vendor contract is included
- Condemned or withdrawn equipment is marked as such and removed from clinical areas
The WHO’s guide on medical equipment inventory management is a helpful reference if you are building the inventory from scratch.
2. Calibration plan and schedule
- A written policy stating which equipment requires calibration and how intervals are decided
- Intervals based on the manufacturer’s recommendation, adjusted for risk and usage
- A calendar or tracker showing due dates for every device on the list
- A process for calibrating new equipment before first use and equipment returning from repair
- A named person responsible for the plan and for chasing overdue items
3. Calibration certificates
- One certificate per instrument, not a single summary sheet for a department
- Serial number and asset ID on the certificate match the device and the inventory
- Results recorded for each parameter, with acceptance limits and a pass or fail statement
- Reference equipment and its traceability stated
- Certificates are current, with no gaps between the last due date and the latest calibration
- If a NABL-accredited lab was used, the parameters are within its accredited scope and the certificate carries the NABL symbol
4. Electrical safety testing
- Mains-powered patient-connected equipment has a recorded electrical safety test
- Results cover protective earth resistance, insulation resistance and leakage current, as applicable to the device class
- Tests are repeated after repair, as well as at the planned interval
Our guide to electrical safety testing of medical equipment explains these tests in more detail.
5. Labelling at the point of use
- A calibration label on each device showing the date done and the next due date
- Labels match the certificate on file
- Out-of-service or failed equipment is clearly tagged so it is not used by mistake
6. Preventive and breakdown maintenance
- A preventive maintenance schedule separate from, but linked to, the calibration schedule
- Breakdown calls logged with date, fault, action taken and downtime
- Equipment history card or log for each critical device
- AMC or CMC contracts on file, with the scope of each clearly stated
7. Handling failures
- A documented process for what happens when a device fails calibration
- Evidence that failed devices were withdrawn, repaired and re-calibrated before returning to use
- Where a device was found out of tolerance, a review of whether it affected patient care during the period it was in use
8. Staff and user training
- Users trained on the safe operation of the equipment they use
- Training records available for critical equipment such as ventilators, defibrillators and infusion pumps
- Staff know how to report a suspected fault and who to contact
Common gaps found before NABH assessments
These are the issues that tend to come up when hospitals review their own files:
| Gap | Why it matters | How to fix it |
|---|---|---|
| Inventory and certificates do not match | Devices appear to be missing calibration | Reconcile serial numbers and asset IDs before the visit |
| Expired certificates | Shows the schedule is not being followed | Use a tracker with alerts well before due dates |
| Department-level certificates | Cannot prove each device was tested | Ask for one certificate per instrument |
| No electrical safety results | Electrical safety is part of the FMS chapter | Add safety testing to each calibration visit |
| No record after repair | Repaired device may not be within limits | Make re-calibration a step in closing a repair job |
| Equipment bought after the last cycle | New devices are left out | Calibrate at installation and add to the tracker |
Does NABH require NABL-accredited calibration?
The NABH 6th edition hospital standard does not spell out that calibration must be done by a NABL-accredited lab, so check the current standard and your assessor’s guidance for your specific case. What is clear is that NABL accreditation gives independent evidence that the lab is technically competent and its measurements are traceable, which makes your records easier to defend. Our article on NABL vs non-NABL calibration covers this in more detail, and how to choose a NABL calibration lab explains what to check before you appoint one.
This checklist is general guidance. Calibration, repair and safety testing should follow the manufacturer’s service manual and be carried out by qualified biomedical engineers.
How Shiwi supports NABH documentation
Shiwi Enterprises is a NABL-accredited calibration lab (ISO/IEC 17025) in Jaipur, working with both NABH-accredited hospitals and those preparing for accreditation. We calibrate on site across India, test each device for performance and electrical safety against calibrated reference analysers, and issue a separate calibration certificate for every instrument. You can see the equipment we calibrate, or read more about our NABL lab.
Get your calibration records assessment-ready
If your NABH assessment or renewal is coming up, share your equipment list with us and we will plan an on-site calibration visit that covers every device and leaves you with certificates ready to file. Request a quote or contact our team to get started.
Frequently asked questions
Which is the current NABH standard for hospitals?
The NABH Accreditation Standards for Hospitals, 6th Edition, effective from 1 January 2025. It has ten chapters, including Facility Management and Safety (FMS).
What calibration records should a hospital keep for NABH?
A complete equipment inventory, a documented calibration schedule, one calibration certificate per instrument, electrical safety test results, maintenance and repair logs, and evidence of how failed devices were handled.
Is electrical safety covered by the NABH standard?
Yes. Electrical safety sits within the Facility Management and Safety (FMS) chapter, so assessors usually ask to see electrical safety test records for patient-connected equipment alongside calibration certificates.
Does NABH require calibration by a NABL-accredited lab?
Check the current standard and your assessor's guidance. NABL accreditation gives independent evidence of a lab's competence and measurement traceability, which makes calibration records easier to defend.
Request a calibration quote
List your equipment with make, model and quantity, and tell us your city. We reply with a quote and a visit plan.
Keep reading
All articles-
Compliance NABL vs Non-NABL Calibration: What Hospitals Should Know
How NABL-accredited and non-accredited calibration differ in scope, traceability and audit value, and how hospitals can tell which kind of certificate they hold.
-
Calibration What Is Medical Equipment Calibration and How Often Should It Be Done?
What medical equipment calibration involves, which devices need it, and how to set calibration intervals based on manufacturer guidance, risk, usage and calibration history.
-
Calibration How to Choose a NABL Calibration Lab for Your Hospital
A step-by-step guide for hospitals on verifying a NABL calibration lab, reading its scope of accreditation, and checking on-site capability, reference equipment and certificates.