NABL vs Non-NABL Calibration: What Hospitals Should Know
How NABL-accredited and non-accredited calibration differ in scope, traceability and audit value, and how hospitals can tell which kind of certificate they hold.

NABL calibration is calibration carried out by a laboratory that NABL has assessed and accredited against ISO/IEC 17025, for parameters within that lab’s accredited scope. Non-NABL calibration may be technically sound, but nobody independent has confirmed the lab’s competence, methods or traceability. For hospitals, the practical difference is the strength of the evidence: a NABL certificate is much easier to defend in an audit or an incident review.
This article explains what separates the two, when each might be used, and what to watch for on the certificates you receive.
What makes a calibration “NABL”?
Three conditions need to be met together:
- The lab holds a current NABL accreditation for calibration. NABL, the National Accreditation Board for Testing and Calibration Laboratories, provides voluntary accreditation to calibration laboratories in accordance with ISO/IEC 17025.
- The parameter being calibrated, and its range, are within the lab’s scope of accreditation.
- The certificate is issued under that accreditation, which normally means it carries the NABL symbol for the accredited parameters.
This last point is often missed. A NABL-accredited lab can still issue a non-accredited certificate if the work falls outside its scope. So the question to ask is not only “is the lab NABL accredited?” but “is this calibration covered by the accreditation?”
How NABL and non-NABL calibration compare
| Aspect | NABL-accredited calibration | Non-NABL calibration |
|---|---|---|
| Independent assessment | Lab assessed by NABL against ISO/IEC 17025 | No independent assessment of competence |
| Scope | Published scope lists accredited parameters, ranges and uncertainties | Lab’s own claims, not independently verified |
| Traceability | Traceability of reference equipment is examined as part of accreditation | May be traceable, but you have to check the evidence yourself |
| Measurement uncertainty | Estimated and reported according to NABL requirements | Sometimes stated, sometimes left out |
| Verification | Lab can be looked up on the NABL website | No central register to check |
| Acceptance in audits | Generally easier to defend | May be questioned, depending on the assessor and context |
| International recognition | NABL is a signatory to the APAC and Global Accreditation Cooperation (formerly ILAC) mutual recognition arrangements | None |
Why traceability is the heart of the difference
A calibration is only as reliable as the reference it is compared against. If a defibrillator analyser itself reads high, every defibrillator tested with it will appear to deliver less energy than it really does.
Metrological traceability means the reference equipment has itself been calibrated, in an unbroken chain, back to national or international standards. NABL guidance (see nabl-india.org) links this to the International System of Units (SI), with equipment calibrated by NPL-India or an accredited laboratory.
With a NABL lab, the assessors have already examined that chain. With a non-accredited provider, the hospital has to ask for and judge the evidence itself, and most hospitals are not set up to do that for every vendor.
Is NABL calibration mandatory for hospitals?
NABL accreditation itself is voluntary for laboratories. Whether your hospital must use NABL-accredited calibration depends on your accreditation body, state requirements, tender conditions or internal quality policy. The NABH 6th edition hospital standard does not spell out that calibration must be done by a NABL-accredited lab, so check the current standard and your assessor’s guidance.
That said, many hospitals choose NABL-accredited calibration for critical equipment because it removes a common point of argument during assessments. If a calibration result is ever questioned after an adverse event, an accredited certificate also gives you a stronger record. Our NABH calibration requirements checklist covers the wider documentation an assessor will look for.
When might non-NABL calibration be acceptable?
There are situations where non-accredited checks still have value:
- Intermediate checks between calibrations. In-house biomedical teams often do quick functional checks, such as confirming a monitor reads a simulator correctly, between formal calibrations. These are useful for catching drift early but do not replace the scheduled calibration.
- Parameters with no accredited provider. For some unusual equipment, an accredited scope may not exist. In that case, ask for evidence of traceability and a clear method.
- Low-risk, non-clinical instruments. Your internal risk assessment may allow simpler verification for equipment that has no effect on patient care.
In each case, document why that approach was chosen. An assessor is more likely to accept a reasoned decision than an unexplained gap.
How to tell what kind of certificate you have
Look at the certificate itself, not the covering letter or the vendor’s brochure:
- Does it show the NABL symbol and the lab’s NABL certificate number?
- Can you find the lab, and that certificate number, in the NABL directory at nabl-india.org?
- Are the parameters on the certificate within the lab’s scope of accreditation?
- Are results, acceptance limits and measurement uncertainty given for each test point?
- Is the reference equipment named, with its own calibration details?
NABL’s guidance to customers is to check for the NABL symbol on reports that cover accredited parameters and to ask the lab for the reason if it has not been used. For a full set of checks before you appoint a provider, see how to choose a NABL calibration lab.
What if your current certificates are non-NABL?
There is no need to discard existing records. They still show that the equipment was checked. A sensible way to move over:
- Sort your equipment by risk. Start with devices that deliver energy, drugs or breathing support, such as defibrillators, infusion pumps, ventilators and electrosurgical units.
- Check due dates. Move each device to accredited calibration at its next due date rather than all at once, unless an audit is close.
- Confirm the new lab’s scope covers each device type before you schedule the visit.
- Keep both sets of records against the asset ID so the history of each device stays complete.
- Update your calibration policy to state which equipment requires accredited calibration and why.
What about electrical safety testing?
Calibration checks whether a device measures or delivers accurately. Electrical safety testing checks whether it is safe to connect to a patient. Both matter, and a complete visit should cover both: leakage current, protective earth resistance, insulation resistance and supply voltage, as appropriate to the device. Our guide to electrical safety testing of medical equipment explains these tests.
This article is general guidance. Calibration and safety testing should follow the manufacturer’s service manual and be carried out by qualified biomedical engineers using suitable reference equipment.
Shiwi’s approach
Shiwi Enterprises is a NABL-accredited calibration lab (ISO/IEC 17025) in Jaipur. We work with NABL-accredited and non-accredited hospitals alike, calibrating on site across India using calibrated reference analysers. Each instrument is checked for performance parameters and electrical safety, and receives its own calibration certificate for NABH and internal audits. You can see the full list of equipment we calibrate, from ventilators to defibrillators.
Move your critical equipment to accredited calibration
If you want to move critical equipment to accredited calibration, we can check your equipment list against our accreditation scope and plan an on-site visit. Learn more about our calibration service, or request a quote.
Frequently asked questions
What is NABL calibration?
Calibration done by a laboratory accredited by NABL against ISO/IEC 17025, for parameters within that lab's accredited scope, with the certificate issued under that accreditation.
Is NABL accreditation mandatory?
NABL accreditation is voluntary for laboratories. Whether a hospital must use NABL-accredited calibration depends on its accreditation body, tender conditions, state rules or internal policy.
Can a NABL-accredited lab issue a non-NABL certificate?
Yes. If the parameter or range is outside the lab's scope of accreditation, the calibration is not covered by the accreditation even though the lab itself is accredited.
How can I tell if my certificate is NABL accredited?
Check for the NABL symbol and certificate number, confirm the lab in the NABL directory at nabl-india.org, and check that the parameters fall within its scope of accreditation.
Request a calibration quote
List your equipment with make, model and quantity, and tell us your city. We reply with a quote and a visit plan.
Keep reading
All articles-
Compliance NABH Medical Equipment Calibration Requirements: A Checklist for Hospitals
A practical checklist for NABH coordinators and biomedical engineers covering equipment inventory, calibration schedules, certificates, electrical safety and the records assessors ask for.
-
Calibration What Is Medical Equipment Calibration and How Often Should It Be Done?
What medical equipment calibration involves, which devices need it, and how to set calibration intervals based on manufacturer guidance, risk, usage and calibration history.
-
Calibration How to Choose a NABL Calibration Lab for Your Hospital
A step-by-step guide for hospitals on verifying a NABL calibration lab, reading its scope of accreditation, and checking on-site capability, reference equipment and certificates.